Portable Three Phase AC Voltage and Current Source GF303D GFuve

The model GF303D three phase power source is electric AC voltage, current and power calibrator for calibration of power meters, power transducers, power quality analyzer, current meter, voltage meter and generally all kinds of power measuring devices.

Description

The model GF303D three phase power source is electric AC voltage, current and power calibrator for calibration of power meters, power transducers, power quality analyzer, current meter, voltage meter and generally all kinds of power measuring devices. AC voltage, current, power functions have calibrated phase shift in frequency range from 40 Hz to 70Hz with resolution as good as 0.001 Hz. It have three channel voltage and three phase current independent output, any programmable setting, low phase shift, high stability 0.01%/1min, as standard three phase voltage source, three phase current source, three phase power source and harmonic source, is a high precision portable calibration tool in electrical laboratory. It can set 3P4W, 3P3W, 1P3W, 1P2W type, forward and reverse outputs, active power and reactive power output.

FEATURES

  • To set up and take the load regulation of voltage, current, phase angle ,frequency and power factor etc; as a large current source, voltage source and power source
  • To set 2 ~ 63 times of the amplitude and phase harmonics, and it can be added to the base wave in every harmonic output
  • Voltage, current output a wide range, big power, high stability, waveform distortion degree is small
  • Strong with load ability, but it can take capacitive, sensibility, resistance of composite type load or load and load regulation is higher than 0.01%
  • To apply the 32 bit MPU + DSP + CPLD, powerful flexible
  • To use hardware PID, fast response, the change of load will not cause output volatility
  • Industrial frequency waves as high as 5000 points a week of the wave kneading, signal output without filter, precise output waveform, the harmonic output precision, harmonic distortion degree is small
  • Range switching automatically
  • To use software calibration, simple operation, stable and reliable
  • Display of vector diagram, adjustable to IEC387 OR DIN410
  • Big touch screen, 7-inch TFT color LCD display, English interface, operating simply
  • To have perfect over-current, over-voltage, heat, a shorts-and-opens, overload protection, failure detection automatically
  • With RS232 interface, and PC connection
  • To take place machine software, it can be output through the PC software program

SELECTION GUIDE

NO. Accuracy Voltage range Current Range Weight
303D12001 0.10% 0-500V 0-120A 25KG
303D12005 0.05% 0-500V 0-120A 25KG
303D1201 0.10% 0-380V 0-12A 18KG
303D1205 0.05% 0-380V 0-12A 18KG
303D2401 0.10% 0-500V 0-24A 20KG
303D2405 0.05% 0-500V 0-24A 20KG

Additional information

Weight 15000 g

Specification

Electrical parameters
Accuracy class 0.02%, 0.05%, 0.1%
Power supply Single phase AC 85-265 V, frequency 50/60 Hz
AC Voltage output
Range (U1, U2, U3 phase) 57.7V/100V/220V/380V; range switch automatically(Max 500V)
Adjust fineness 0.01% RG
Accuracy 0.02% RG, 0.05% RG
Stability <0.01% RG/120s
Distortion degree <0.3% (not capacitive load)
Output power 25VA, 50VA or 100VA
Full load regulation rate 0.01% RG
Full load regulation time Less than 1mS
Long-term stability ± 60 PPM/year
AC Current output
Range (I1, I2, I3 phase) 0.2A, 1A, 5A, 20A, 100A; range switch automatically
Adjustment range (0-120)%RG
Adjust fineness 0.01% RG
Accuracy 0.02% RG, 0.05% RG
Stability <0.01% RG/120s
Distortion degree <0.3% (not capacitive load)
Output power 50VA, 100VA or 200VA
Full load regulation rate 0.01% RG
Full load regulation time Less than 1mS
Long-term stability ± 60 PPM/year
Power output
Accuracy 0.02% RG, 0.05% RG
Stability 0.01% RG/120s
Phase angle
Adjusting range 0.000°-359.999°
Output adjustment fineness 10, 1, 0.1, 0.01 as optional.
Resolution 0.001°
Accuracy 0.02° or 0.05°
Power factor
Adjusting range -1 ~ 0 ~ +1
Resolution 0.0001
Accuracy 0.0005
Frequency
Adjusting range 40.000-70.000 Hz
Output adjustment fineness 5Hz, 1Hz, 0.1Hz, 0.01Hz as optional.
Resolution 0.001 Hz
Accuracy 0.002Hz
Temperature drift ± 0.5 PPM/°C
Long-term stability ± 4 PPM/year
Harmonic accuracy
Harmonic number 2-63times
Harmonic content 0-40%
Harmonic phase 0°-359.99°
Harmonic phase accuracy < 0.01°
Harmonic set accuracy 0.1% (relative to the base wave ratings)
Functions
Communication Port RS232
Programmable controlled Yes
Harmonic Yes
Wiring mode 3P4W, 3P3W, 1P3W, 1P2W
Key 30pcs
LCD 7 inch TFT color touch display
PC control software Optional
Standard
Standard IEC 62053-21,22, 23; IEC 60736; ANSI C12.20-2002;
JJG 597-2005; JJG596-2012; JJG 1085-2013; JJF 68-2019;
DL/T 826-2002; DL/T 1478-2015; DL/T 448-2016; JJG 51-1999;
DL/T 2213.1-2021; GJB J 5857-2006; JJG6-2011; JJG70-2015;
Safety
Isolation protection IEC 61010-1:2001
Measurement Category 300 V CAT III, 600 V CAT II
Degree of protection IP20
Declaration of conformity CE & CNAS certified
Mechanical parameters
Dimensions (W×D×H) (mm) 500x600x175
Weight (kg) 12, 15, 18, 20, 25
Environmental conditions
Working temperature 0°C to 50°C
Storage condition -30°C to -60°C
Relative humidity ≤ 85%

Downloads

Manual

Request For Quotation

Enjoy Special Offer Only For You

Contact Us :