TG Flue gas temperature 4°F – 2192°F 0°C – 1200°C
TA Air temperature 4°F – 248°F 0°C – 120°C
O2 Oxygen0 20 9 Vol %
CO Carbon Monoxide Electrochemical up to 100 000 ppm
CO Carbon Monoxide NDIR 0 1020 Vol %
CO2 Carbon Dioxide Calculated 0 CO2 max
CO2 Carbon Dioxide NDIR 0 20 Vol %
NO Nitric Oxide 0 2000 ppm
NO2 Nitrogen Dioxide 0 100 ppm
NOx Nitrogen Oxides 0 NOx max
SO2 Sulfur Dioxide 0 4000 ppm
H2S Hydrogen Sulfide 0 200 ppm
HCCH4 Hydrocarbons 0 100% LEL
NH3 Ammonia 0 10005000 ppm
N2O Nitrous Oxide 0 1 Vol %
HCl Hydrogen Chloride 0 200 ppm
Cl2 Chlorine 0 5000 ppm
H2 Hydrogen 0 10000 ppm
Losses Efficiency Calculation 0 99 9 %
Excess Air Lambda Calculation 1 0 9 99
P Draft + 40 hPa
Velocity with Pitot tube 0 80 ms
Soot Filter paper method 0 9
Other ranges are available upon request
IMR Environmental Equipment Inc reserves the right to adopt technical modifications without prior notice
Ini hanya demo
Free
Ini hanya demo saja
2-3 Days
Free
The IMR 2800P is a portable Flue Gas Analyzer for industrial applications
The rugged combustion analyzer can be equipped with up to 8 sensors
Its basic unit has an O2 Oxygen sensor a CO Carbon Monoxide sensor a NO Nitric Oxide sensor and a SO2 Sulfur Dioxide Sensor The other sensors are customer select able as well as the measuring ranges
The IMR 2800P is a portable continuous measuring flue gas analysis instrument housed in an aluminum case
It measures flue gas temperature air temperature Oxygen O2 Carbon monoxide CO Nitric Oxide NO and Sulfur Dioxide SO2
It can optionally measure Nitrogen Dioxide NO2 Sulfur Dioxide SO2 Hydrogen Sulfide Hydrocarbons HC and much more
Combustible calculations like Losses Excess Air and Carbon dioxide CO2 are standard as well as the measurements for Draft and optionally Soot
Further features include
23 Fuels are programmed – 4 fuels are free programmable
Automatic zero calibration
Integrated self check program
Simultaneous display parameters on the illuminated display
Printer with programmable print out cycles
RS232 DB9USB
Standard deviation and average value calculation
Draft measurement with optional upgrade for differential draft and velocity measured with a Pitot tube
Unit selection ppm – mgNm3– mgNm3 ref O2
Gas sampling probe E – length 0 8 ft 250mm hose 8 ft 2 5m
Rechargeable battery with charger
Power supply 110V or 230V or 12VDC Power Jack