KS813 Three phase AC and DC Calibrator

Output Measuring AC Voltage

Available Ranges

10V 30V 100V 300V 750V auto shift

Adjustment range

0~120%Rg

Adjustment fineness

Range x 0 01%

Resolution

Range x 0 01%

Accuracy

0 05%Rg

Stability

0 01% 1min

Output Measuring AC Current

Available Ranges

100mA 1A 5A 10A 25A; auto shift

Adjustment range

0~120%Rg

Adjustment fineness

Range x 0 01%

Resolution

Range x 0 01%

Accuracy

0 05%Rg

Stability

0 01% 1min

Output Measuring AC Power

Adjustment fineness

Range x 0 01%

Resolution

Range x 0 01%

Accuracy

0 05%Rg Factor>0 5

Stability

0 01% 1min

Output Measuring AC Frequency

Frequency Range

45 000~65 000Hz

Adjustment fineness

0 001Hz

Accuracy

0 01%RD

Output Measuring AC Phase Angle

Phase Angle Range

0 00º~359 99º

Adjustment fineness

0 01º

Resolution

0 01

Accuracy

0 05º

Output Measuring AC Power Factor

Output Range

1 ~ 0 ~ +1

Measuring Accuracy

0 0005

Adjustment fineness

0 0001

Output Measuring Harmonic

Harmonic Setting

2~31

Harmonic Content

voltage current ≤30% as against fundamental

Harmonic Output Accuracy

0 1% 1st~19th as against fundamental

0 2% 20th~31st as against fundamental

Harmonic Phase

0~360º adjustable

Measuring Electrical Energy

Measuring Accuracy

0 1%Rd PF≥0 5

Voltage Range

100V 220V 380V

Current Range

0 05~24A

Distortion of AC Outputs

Distortion

<0 2% non capacitance load Max Load Capacity of AC Voltage Current Voltage Output 25VA max Current Output 25VA max Output Measuring DC voltage Range 100mV 1V 10V 30V 100V 300V 750V Adjustment Range 0~120%Rg 0%~110% output range at 750V Resolution Range x 0 01% Accuracy 0 05%Rg Stability 0 01% 1min Output Measuring DC current Range 1mA 10mA 100mA 1A 5A 10A 25A Adjustment Range 0~120%Rg Adjustment fineness Range x 0 01% Resolution Range x 0 01% Accuracy 0 05%Rg Stability 0 01% 1min Max Load Capacity of DC Voltage Current Voltage Output 20W max Current Output 25W max Power Supply Environment Ambient Temperature 22ºC ± 1ºC Working Temperature 0ºC~40ºC humidity≤85%RH Nominal Input Voltage 115230Vac appointed ±10% Nominal Frequency 5060Hz Connection Type Standard AC socket 60320 Weight Size Dimensions W x D x H 450mm×380mm×190mm Weight 25kg

12 People watching this product now!
  • Ini demo, bisa diisi apa saja
  • Ini demo, bisa diisi apa saja
  • Ini demo, bisa diisi apa saja
  • Ini demo, bisa diisi apa saja
  • Produk Tridinamika tersertifikasi jaminan SNI

Ini hanya demo

Free

  • Jasa Pengiriman kilat oleh tim Tridinamika

Ini hanya demo saja

2-3 Days

Free

Product Details

Specification

Description

KS813 Standard Source Transducer Calibrator Three phase ACDCHarmonic Standard Source is a versatile instrument suitable for a wide range of applications in power companies power plants research institutions and universities It features open protocol for software self programming and is fully compatible with national standards The AC voltage sources provide 3x900V@3x25VA output with 0 05% accuracy while the AC current sources deliver 3x30A@3x25VA output with 0 05% accuracy The instrument also offers 3 phase Harmonic Standard Power with 2 31x orders and 0 1% accuracy DC voltage sources with 825V@20W max and 0 05% accuracy DC current sources with 30A@25W max and 0 05% accuracy The KS813 comes with a 3 year guarantee of free repair and life long maintenance as well as free software updates and training The instrument has a storage capacity of up to 1 000 calibrated meters’ data and its 5 4” TFT LCD display touch pad and rotary encoder control make it easy to use

KS813 Standard Source Transducer Calibrator also has an open communications protocol that facilitates the secondary development of various factory automatic verification devices

Additional information

Additional information

Brand

Kingsine

Contact Us