MS1900 – Oil in Water Analyzer

Detection Range Minimum 0 1 ppm maximum 20 ppm

Output Range Default 0 ppm to 20 ppm configurable during commissioning

Repeatability 2% to +2% see Note 1

Accuracy 15% to +15% see Note 1

Measurement Frequency Continuous measurement with optional modes of 15 20 30 and 60 minutes available upon request

Ambient Temperature 0 °C 32 °F to 40 °C 104 °F see Note 2

Water Temperature 1 °C 34 °F to 40 °C 104 °F see Note 2

Analogue Output 4 mA to 20 mA scalable to the required range with a maximum load of 900 Ω

Analogue Output Isolation 400 V DC

Relay Voltage Maximum 50 V with alarm and fault relays featuring NO contacts

Relay Current Maximum 5 A

Service Interface USB A to PC using the provided Multisensor Software

Data Storage Uses a μSD card with up to 6 months of storage capacity

Instrument Case Rated IP65 NEMA 4X

Instrument Weight 5 kg 11 lbs

Instrument Dimensions 300 x 200 x 132 mm 11 8 x 7 8 x 5 2 inches

Sampling Chamber Material Stainless steel with PVC available as an option

Sampling System Weight 12 kg 26 lbs without water

Sampling System Dimensions 570 x 490 mm 22 x 19 inches mounted on its own PVC backboards

Sampling System Capacity 3 liters 0 79 US gallons

Flow Limit Switch Contacts close if flow falls below the set point option available upon request

Water Flow Rate 2 lmin 0 52 US gpm

Supply Voltage

AC Version 90 V AC to 240 V AC 50 Hz or 60 Hz
DC Version 10 V DC to 15 V DC
Power Consumption

Maximum 9 W
Typical during operation 7 W
Note 1 Accuracy measured against a 10 ppm toluene solution at standard temperature and pressure Accuracy can be further improved at points of interest using bespoke calibration

Note 2 Climate control may be required depending on process and ambient conditions

12 People watching this product now!
  • Ini demo, bisa diisi apa saja
  • Ini demo, bisa diisi apa saja
  • Ini demo, bisa diisi apa saja
  • Ini demo, bisa diisi apa saja
  • Produk Tridinamika tersertifikasi jaminan SNI

Ini hanya demo

Free

  • Jasa Pengiriman kilat oleh tim Tridinamika

Ini hanya demo saja

2-3 Days

Free

Product Details

Specification

The MS1900 is an on line Oil in Water analyzer designed to provide continuous measurement of VOCs in industrial and trade effluent discharge points Regulations on the quality of effluents from trade premises and factories are getting stricter and stricter and VOCs are now in the spotlight

Applications
Monitor water abstraction points such as raw river water boreholes and reservoir water
Monitoring for oil in water contamination
Detect VOC contamination in industrial wastewater
Monitor drain and stormwater systems for flammable liquids
Surface water systems for fuel pollution
Offer a monitoring solution for air land remediation and industrial applications
Detect VOC breakthrough in carbon beds
RO membrane protection
Measurement of VOCs in process water

Features and Benefits
No sensor contact with water low maintenance no sensor cleaning
No reagents low running costs
Measures both Polycyclic Aromatic Hydrocarbons PAHs and VOCs
Electronic Nose not affected by turbidity
High sensitivity ideal for potable water
Can be calibrated for specific substances used to monitor known spills
Catches VOC chemical spills not just hydrocarbons
Alarms to SCADA and other communication interfaces
For more information on the principle of operation click on the FAQs tab

Request a Quote

Industrial Effluent The Problem
Discharge of VOCs into wastewater is an endemic problem in engineering chemical processing food and beverage production and other industrial processes

Spills can lead to illegal discharges pollution fines and PR disasters

For this reason some industrial facilities have purification plants to remove the bulk of these contaminants from the water before it is discharged but even these suffer from failures and can be over run

The cost of not monitoring your effluent
When spills happen large fines can be issued by the Environmental Agency and local national and social media can turn an accident into a PR disaster with long term implications

Additionally it is likely that regulatory bodies will ask for expensive on line TOC analyzers to be installed which require a lot of reagents and maintenance at a very high additional cost

The Solution
By installing a VOC Industrial Effluent Discharge Monitor at the effluent you can be notified as soon as the values start to increase and take appropriate action avoiding all the problems associated with unexpected pollution events

The VOC Industrial Effluent Discharge Monitor is used to check the industrial effluent from industrial facilities and business premises to ensure that it is complying with environmental regulations

Its contactless measurement technique means that

It does not need reagents
Requires little or no maintenance between services
It’s robust and reliable
It is cost effective
Thanks to this analyzer you can monitor your effluent discharge and make sure that you detect promptly any problem avoiding expensive fines and PR disasters The system comprises the analyzer and the sampling system As the water goes through the sampling tank VOCs dissolved in the water move into the gas phase

Multisensor’s sampling system helps the VOCs to get in the gas phase This gas is then continuously analyzed and the VOCs concentration in the water is determined

For applications where drains are being monitored the MS1800 can be used to directly sample gases on the drain This provides a lower cost solution but reduces absolute accuracy